Hach China
400-686-8899 / 800-840-6026
 

Displaying RGT/SMP Error on the CA610 Fluoride analyzer

产品分类:CA610在线氟化物分析仪
更新日期:2018-11-09
● 问题咨询

Displaying RGT/SMP Error on the CA610 Fluoride analyzer

● 关键词摘要

CA610,在线氟化物

● 解答

RGT/SMP Error is on software Ver 1.4 and can be cleared on the CA610 Fluoride analyzer.  It may be caused by bubbles in the sample stream, or that cling inside the measurement chamber.  The message doesn't happen as often with the latest Ver 1.5 software.  If this is a reoccurring problem, and the instrument has Ver 1.4, the entire circuit board has to be replaced in order for it to update to the newest version.